Results of RhD testing by the tube and plate methods using a serially diluted anti-D reagent
Dilution ratio | Reactivity | Time to agglutination | ||
---|---|---|---|---|
Tube | Plate | Plate | ||
1X | 4+ | 3+ | 10 s | |
2X | 4+ | 3+ | 10 s | |
4X | 4+ | 3+ | 10 s | |
8X | 4+ | 3+ | 10 s | |
16X | 4+ | 3+ | 13 s | |
32X | 4+ | 2+ | 17 s | |
64X | 4+ | 2+ | 26 s | |
128X | 4+ | 2+ | 30 s | |
256X | 3+ | 2+ | 45 s |
These are representative data from at least three experiments. A dilution ratio of 1 indicates that the original anti-D reagent was undiluted.