Cases of adverse events associated with blood transfusion-related medical devices (2013∼2022)

Case Device Model Medical device problem* Clinical signs, symptoms and conditions* Health impact* Year of report
1 LRF NA (Haemonetics corporation, MA, USA) A1408_ No flow E2403_No clinical signs, symptoms or conditions F05_Delay to treatment/ therapy
F2301_Additional device required
2013
2 LRF NA (Haemonetics corporation, MA, USA) A050401_Fluid leak E2403_No clinical signs, symptoms or conditions F1004_Underdose 2013
3 LRF NA (Haemonetics corporation, MA, USA) A140902_Partial blockage E2403_No clinical signs, symptoms or conditions F05_Delay to treatment/ therapy
F2301_Additional device required
2014
4 BTS NA (Doowon meditec corporation, Korea) A050401_Fluid leak E2403_No clinical signs, symptoms or conditions F2301_Additional device required 2015
5 LRF NA (Haemonetics corporation, MA, USA) A1408_ No flow E2403_No clinical signs, symptoms or conditions F2301_Additional device required 2015
6 LRF NA (Haemonetics corporation, MA, USA) A1408_ No flow E2403_No clinical signs, symptoms or conditions F2301_Additional device required 2015
7 LRF NA (Haemonetics corporation, MA, USA) A050401_Fluid leak E2403_No clinical signs, symptoms or conditions F2301_Additional device required 2016
8 LRF RC2VAE (Haemonetics corporation, MA, USA) A050401_Fluid leak E2403_No clinical signs, symptoms or conditions F1004_Underdose,
F2301_Additional device required
2016
9 LRF RC2VAE (Haemonetics corporation, MA, USA) A040101_Fracture E2403_No clinical signs, symptoms or conditions F2301_Additional device required 2018
10 BTS 903-00004 (Belmont instrument corporation, CA, USA) A050401_Fluid leak E2403_No clinical signs, symptoms or conditions F2301_Additional device required 2021
11 BTS 903-00006 (Belmont instrument corporation, CA, USA) A040101_Fracture E2403_No clinical signs, symptoms or conditions F2301_Additional device required 2021
12 LRF RC2VAE (Haemonetics corporation, MA, USA) A1408_ No flow E2403_No clinical signs, symptoms or conditions F2301_Additional device required 2022

*International Medical Device Regulators Forum codes were used to classification (code_term); A1408_ No flow, A050401_Fluid leak, A140902_Partial blockage, A040101_Fracture, E2403_No clinical signs, symptoms or conditions, F05_Delay to treatment/ therapy, F2301_Additional device required, F1004_Underdose.

Abbreviations: LRF, leukocyte reduction filter; NA, not available; BTS, blood transfusion set.

Korean J Blood Transfus 2022;33:161-70 https://doi.org/10.17945/kjbt.2022.33.3.161